The new iNEXIV VMA-2520 has been expressly designed for three-dimensional workpieces. Though compact and lightweight in design, it provides a long 200mm Z-axis stroke and a 73.5mm working distance, enabling easy Z-axis measurement of mechanical parts, plastic injection molding parts, metal press parts, electronic devices, dies & molds and medical devices. The standard 10x zoom optics meets the industry’s demanding needs for superb resolution at high magnifications while offering a wide field of view at low magnifications. Low distortion optics and high intensity white LED illumination sources improve contrast to enhance throughput. This combination assures reproducible measurements even for colorful parts.


Main features


Easy and accurate measurements optimized for 3D parts

Nikon has drawn on its state-of-the-art digital technologies to produce a highly affordable automated measurement system. The iNEXIV VMA-2520’s design is optimized for easy use as well as repeatable and accurate measurement of 3D parts.

Compact and lightweight

The iNEXIV VMA-2520 is a powerful bench top system designed to use minimum factory floor space. The main body has a footprint of only 565mm W x 690mm D x 740mm H (computer and controller separate) and weighs only 72kg (158.7 lb), yet it still has a 250mm x 200mm x 200mm measuring capability.

Long working distance

The newly developed optical system features a super long 73.5mm working distance for all magnifications. This allows sufficient three dimensional clearance for Z-axis measurements even at high magnification.

Sophisticated VMA AutoMeasure software

This new versatile software has been developed exclusively for the VMA-2520. It provides various functions to easily conduct tasks ranging from setup, teaching programs and measurements, to evaluations.

High-speed, highly-accurate Laser AF (optional)

In addition to the standard Vision AF, an optional high-speed Laser AF with a long 63mm working distance is available. It provides fast focusing and Z-axis measurements regardless of the surface shape of your workpiece.

Touch probing measurement (optional)

The iNEXIV VMA-2520 accepts the Renishaw TP20 system, making 3D parts measurement much easier.

  • Touch Probing Option will be available in 2007 summer.

Measurement Examples

Electron gun
Electron gun
Metal press parts
Metal press parts
Connector
Connector
Plastic injection molding parts
Plastic injection molding parts
Aluminum die casting parts
Aluminum die casting parts
Drill
Drill

Specifications

iNEXIV VMA-2520 main body
Stroke (X x Y x Z) 250 x 200 x 200 mm (10"x 8" x 8")
Minimum readout 0.1μm
Maximum workpiece weight 15kg (up to 5kg accuracy guaranteed)
MPE* XY MPEE1: 2+8L/1000μm
XY MPEE2: 3+8L/1000μm
Z MPEE1: 3+L/50μm
Camera 1/3-in. 3CCD color
Progressive scan (B/W optional)
Working distance 73.5mm (63mm with LAF)
Magnification Optical: 0.35 to 3.5x
On screen: 12 to 120x (with 17" monitor at SXGA resolution)
FOV size 13.3 x 10 to 1.33 x 1mm
Auto focus Vision AF and optional Laser AF
Illumination  
  Contour illumination White LED illumination
  Surface illumination White LED illumination
  Oblique illumination 8-segment white LED ring
Video resolution 640 x 480 (pixels)
Touch probe Renishaw® TP200/TP20 (option)
Power source 100V-240V, 50/60Hz
Power consumption

5A-2.5A
(excluding power consumption of host computer and its peripherals)

Dimensions & weight  
  Main body
(W x D x H)
565 x 690 x 740 mm (minimum height), 72kg
  Controller 145 x 400 x 390 mm, 13kg
  Table for main body (option) 600 x 700 x 825 mm, 38kg
Recommended workbench Min. load capacity 1000kg or more
Operational environment  
  Temperature 10ºC to 35ºC
  Humidity 70% or less
Host computer
CPU Intel Core Duo 1.8GHz or faster
Memory 1G or more
Operating system Windows® XP Pro SP2 or later
Interface USB2.0/IEEE1394

*Nikon's in-house test at 20℃ ±0.5k